我用AFE4400 模拟前端 做一个心率计,发现读出六个寄存器里的值只有 高八位看起来有效,中八位和低八位 要么是0xFF,要么是0x00, 而且指压传感器后,数据变化都不大,请问这正常吗?下面 附一组数据(一秒钟采一次样),供参考:
0F FF FF 5D FF 00 0C FF FF 1A 00 00 0A 00 FF 51 FF FF0F FF FF 5D 00 00 0C 00 FF 19 FF 00 0A FF FF 51 00 FF0F 00 FF 5D FF 00 0C 00 FF 19 00 00 0A FF FF 51 00 FF
0F FF FF 5D FF 00 0C 00 FF 19 FF 00 0A 00 FF 50 00 FF
0F 00 FF 5C 00 00 0B 00 FF 19 FF 00 0A FF FF 50 FF FF0E FF FF 5C 00 00 0B 00 FF 19 FF 00 0A 00 FF 50 FF FF
0E 00 FF 5C FF 00 0B 00 FF 19 FF 00 0A 00 FF 50 00 FF
Van Yang:
不正常,正常情况你下你通过excel将数据画图可以见到明显的PPG波形